Product Category

4153 IN

Datasheet print icon-img

FEATURES

  • Microprocessor-controlled.
  • Measures Insulation Resistance, voltage (AC-DC) measurements with automatic hold facility, diode, ohm, and continuity with a short circuit current of minimum 220mA.
  • Auto-off function.
  • Energy conservation feature.
  • Ener-Save™ limit the test duration to about 10 Seconds.
  • Quick Test and EnerSave™ are unique features. Both saves energy.
  • No moving parts. All calibration are saved internally in a non volatile memory.
  • Easy calibration at any facility around the world.
  • Operates with rechargeable batteries, alkaline or low cost general purpose batteries.
  • Low cost maintenance, calibration, and ownership.
  • Complies to all the latest regulations, including UK.

SPECIFICATIONS

Insulation Test
Test voltage
(DC V)
250V / 0.2M~2GΩ
500V / 0.2M-4GΩ
1000V / 0.2M-8GΩ
Output voltage @ 1mA Rated test Voltage +10% Max
Accuracy 0.2M~4GΩ : ±3%rdg
4GΩ~8GΩ : ±5%rdg
When voltage is constant, Current is limited at 1.2mA
Short circuit current 4mA Max.
Continuity
Measuring ranges 0.01-1999Ω
Accuracy 0.01-100Ω : ±1.0%rdg
100-300Ω : ±1.5%rdg
300-1999Ω : ±2.0%rdg
Short circuit current >220mA
Open circuit voltage 5V dc
Resolution ±2 counts
Voltmeter
AC voltage (Auto) 0-700V
DC voltage (Auto) 0-950V
Accuracy ±1.5% rdg
Resolution ±1V
Protections
Over load 700V
(between all terminals)
Over voltage Class III-700V towards ground
Fuse 500mA 250V, (5 × 20mm), HBC, Fast Blow
General
Display 2 lines × 16 characters LCD
Auto-null threshold
Buzzer threshold
Fast test 10Sec
Long test 60Sec
With PI and DAR test function 10Min
Load battery test current About 300mA
Operating temperature 0°C to 40°C not in full sun
Storage temperature -10°C to 50°C
Dimensions 250(L) × 190(W) × 110(D)mm
Weight
(battery included)
Approx 1460g.
Power source 1.5V(AA) × 8
Safety standard EN 61010-1 CAT III 600V