FEATURES
- The RCB-411 is used to verify and proof High Voltage Insulation Testers.
Both Analogue and Digital Insulation Testers can be calibrated. - Caution:
- Before testing, make sure the front panel, terminals are clean and dry. If not, wipe them with cloth.
- Use high voltage silicone wires.
Otherwise the reading will be incorrect.
SPECIFICATIONS
| Resistance ranges (Ω) | 1MΩ, 10MΩ, 100MΩ, 500MΩ, 1GΩ, 10GΩ, 50GΩ, 100GΩ, 1TΩ |
|---|---|
| Resistance tolerances | ±1%(F) |
| Tolerance | ± 25ppm/°C: 1MΩ, 10MΩ, 100MΩ ± 50ppm/°C: 500MΩ ± 100ppm/°C: 1GΩ ± 200ppm/°C: 10GΩ, 50GΩ, 100GΩ ± 1500ppm/°C: 1TΩ |
| Max. working voltage | DC 5kV |
| Operating temperature | -30°C~70°C |
| Dimensions | 250(L) × 190(W) × 110(D)mm |
| Weight | Approx. 1455g |